Part I Service Scenarios
Maintenance 54
Figure 2-2 Mechanical safety check flow
Check handle of main unit
Check control panel And
turning
mechanism
Check other mechanics
Check the supporting part of
cart
Check
the handle of cart
(optional)
Check the connection of cart
casters
Check caster
s of cart
(optional)
Check enclosure of main unit
Following is the mechanical safety checklist:
No. Item Method Tool
1 Caster 1 Visual inspection: Check that casters do not
crack.
2 Operate the casters to check that the locking
and releasing functions are normal.
None
2 Caster connection 1 Visual inspection: Check that the casters do
not skew, and that the screws do not crack or
fall off.
2 Check with the spanner to make sure that
there is no looseness between the caster and
the base connection screw.
Inner
hexagon
spanner
3 Trolley handle 1 Visual inspection: Check that the handle has no
cracks.
2
Check by hand to ensure that the handle does
not loosen.
Inner
hexagon
spanner
4 Trolley supporting
parts
Use hands to check that the supporting parts do not
loosen.
/

Table of Contents

Other manuals for Mindray MX7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray MX7 and is the answer not in the manual?

Mindray MX7 Specifications

General IconGeneral
Display Size15.6 inches
Transducer Ports3
PortabilityPortable
TypeUltrasound System
Display TypeLCD
Battery Life120 min
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler
ConnectivityWi-Fi, Ethernet, USB
ApplicationsVascular, Cardiology, OB/GYN, Urology
Probe Connectors3
Operating SystemWindows

Related product manuals