1-4 Safety Precautions
Do not use an aftermarket probe other than those specified
by Mindray. The probes may damage the system, causing a
profound failure, e.g. a fire in the worst case.
Do not subject the probes to knocks or drops. Use of a
defective probe may cause an electric shock.
Do not open the covers and front panel of the system.
Short circuit or electric shock may result when the system
hardware is exposed and powered on.
DO NOT use this system simultaneously with equipment
such as an electrosurgical unit, high-frequency therapy
equipment, or a defibrillator, etc. This would result in a risk
of electric shock to the patient.
When moving the system, you should first fold the LCD
display, disconnect the system from other devices
(including probes) and disconnect the system from the
power supply.
Accessory equipment connected to the analog and digital
interfaces must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety
standard and IEC 60601-1 medical equipment standard).
Furthermore, all configurations must comply with the
standard IEC 60601-1 chapter 16 ME System. It is the
responsibility of the person, who connects additional
equipment to the signal input or output ports and
configures a medical system, to verify that the system
complies with the requirements of IEC 60601-1 chapter 16
ME System. If you have any questions regarding these
requirements, consult your sales representative.
Prolonged and repeated use of display controls may result
in hand or arm nerve disorders for some individuals.
Observe the local safety or health regulations concerning
the use.
DO NOT contact both the patient and the ultrasound
system or the live parts of the ultrasound system (e.g.
signal I/O ports). Electric shock may occur.
When using intra-cavity transducers, do not activate the
transducer outside the patient’s body.
If you have any doubts of the installation or routing of
external protective cables, use the internal power supply
of the system.
DO NOT put the ultrasound system in any soft materials
(e.g. soft cloth or cotton, etc.) in case that the air vent
becomes blocked.
Always read and follow carefully the manufacturer
instructions on the contrast agent label.
Only use the ECG leads provided with the ECG module;
otherwise electric shock may result.