Safety Precautions 1-5
Precautions concerning clinical examination techniques:
This system must be used only by qualified medical
professionals.
This operator’s manual does not describe clinical
examination techniques. The clinician should select the
proper examination techniques based on specialized training
and clinical experience.
Malfunctions due to radio waves:
If a device emitting radio waves is used in the
proximity of this system, it may interfere with
operations. DO NOT use or take any devices
transmitting RF signals (such as cellular phones,
transceivers and radio controlled products) into the
room where the system is located.
If a person brings a device that generates radio
waves near the system, ask him / her to immediately
turn OFF the device.
Precautions concerning movement of the system:
When you place the system on the trolley and move
them together, you must secure all objects on the
trolley to prevent them from falling. Otherwise you
should separate the system from the trolley and
move them individually.
When you have to move the system with the trolley
upward or downward the stairs, you must separate
them first and then move them individually.
Object placed on the display may fall and injure an
individual when moving.
Confirm that there is no peripheral device connected
to the system before moving the system. Otherwise,
peripheral device may fall and cause injury.
DO NOT expose the system to excessive vibration through
transportation. Mechanical damage may result.
Do not connect this system to outlets with the same circuit
breakers and fuses that control the current of devices such
as life-support systems. If this system malfunctions and
generates over current, or when there is an instantaneous
current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.
Always keep the system dry. Avoid transporting this system
quickly from a cold place to a warm place; otherwise
condensation or water droplets may form allowing a short
circuit and possible electric shock.
If the circuit protector is tripped, it indicates that the system
or a peripheral device was improperly shut down and that the
system is unstable. You cannot repair the system under this
circumstance and must call the Mindray Customer Service
Department or sales representative.