DICOM/HL7 11-11
Name
Description
Magnification
Type
Select how the printer magnifies an image to fit the film.
Replicate: interpolated pixels belong to duplicates of adjacent
pixels)
Bilinear: interpolated pixels are generated from bilinear
interpolations between adjacent pixels
Cubic: interpolated pixels are generated from cubic
interpolations between adjacent pixels
None: without interpolation.
Add
Add the DICOM service to the service list
Cancel
Select to cancel parameter preset.
Update
Select an item in the service list, change the parameters in the
above area, and tap [Update] to update the item in the service
list.
Service
List
Delete
Select to delete the selected service from the service list.
Default
Select an item in the service list. Tap [Default] and you will see
“Y” in the Default column.
Verify
Select to verify that the two DICOM application entities are
properly connected.
11.1.3.3 DICOM Worklist Preset
1. On the DICOM/HL7 screen, select the [Worklist] page tab to enter the Worklist page.
2. Select a device and enter the correct AE Title, port, etc.
You can preset what DICOM elements that will not be used in worklist query by using
[Remove Attributes( )] button.
3. Tap [Add] to add the service to the Service List.
The DICOM Worklist service parameters are similar to those described in DICOM Storage Preset.
See “11.1.3.1 Storage Service Preset” for details.
11.1.3.4 MPPS Preset
1. On the DICOM/HL7 screen, select the [MPPS] page tab to enter the MPPS page.
2. Select a device and enter the correct AE Title, port, etc.
3. Tap [Add] to add the service to the Service List.
The DICOM MPPS service parameters are similar to those described in DICOM Storage Preset.
See “11.1.3.1 Storage Service Preset” for details.
11.1.3.5 Storage Commitment Preset
1. On the DICOM/HL7 screen, select the [Storage Commitment] page tab to enter the Storage
Commitment page.
2. Select a device and enter the correct AE Title, port, etc.
3. Tap [Add] to add the service to the Service List.
A special setting item for the DICOM Storage Commitment service is the Associated Storage
Service, as described in the following. Other parameters are similar to those described in DICOM
Storage Preset. See “11.1.3.1 Storage Service Preset” for details.