13-40 Probes and Biopsy
4. Hold the transducer. Press the biopsy needle pressure position to separate it from needle
guided V-shaped block.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
6. Hold the transducer, and release the pressure position of the needle. Adjust the needle-type
adjusting nut manually (following the direction of the arrow). The needle moves smoothly at
the vertical direction due to its gravity.
NGB-043
1. Put on the sterile probe sheath.
2. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the other
hand. Match the grooves of the needle-guided bracket to the tabs of the transducer. Mount the
bracket onto the transducer, and rotate the locking nut in the direction of the arrow shown on
the figure to fix the bracket.