EasyManua.ls Logo

Mitutoyo EH Series - 3 Setting Parameters; 3.1 Procedure for Setting Parameters

Mitutoyo EH Series
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Setting Parameters
11
No. 99MBC109B
3
Setting Parameters
The settings of the Linear Gage that you will use, the display of the Counter,
and external output are specied by setting parameters. Set parameters before
you begin measuring.
2-axis models have 4 internal counters that are referred to as "CEL", and
depending on the Display mode setting, 6 types of counter values can be
specied for each CEL separately. For details, see
"3.4 Details of Display
Mode (2-axis Models Only)" (page 17).
3.1
Procedure for Setting Parameters
Parameters are set in Parameter mode. As an example, the procedure for using
the Linear Gage with a resolution of 5 µm for EH-102P is explained.
1
Turn on the power switch.
» The Counter enters the stand-by
state.
2
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter enters Parameter mode.
Parameter number Set value
3
Press [P.SET] repeatedly to advance the parameter number to 12.
» The current value of parameter
number 12 of INPUT A will be
displayed.
(Parameter number 12 sets the
resolution.)
Parameter number
INPUT
number
Set value
4
Repeatedly press [A_ZERO]/[ZERO] to set the set value to 1 (resolution:
5 µm).
» The value of INPUT A will be set to 1
(the Linear Gage resolution of INPUT
A will be set to 5 µm).
Parameter number
INPUT
number
Set value

Related product manuals