3 Setting Parameters
14
No. 99MBC109B
3.3
Advanced Parameters
This section explains the parameters related to the display, functions, and
external output of the Counter. Congure the settings appropriate to your
application.
Tips
The circles in the Counter model columns in the following table indicate whether a
parameter number that is displayed on the Counter is valid or invalid (●: valid, ○:
invalid).
No. Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
01
Parameter initial-
ization
If you set the value of this parameter to 1, the set
values for all parameters, except for the unit setting,
can be reset to their default values (initialized).
Once this setting has been enabled, this parameter is
reset so its set value is 0 (do not initialize).
● ● ● ●
0: Do not initialize 1: Initialize
02
Key protect Key operations can be disabled to prevent operation
errors.
● ● ● ●
0: Key operation enabled
1: Key operation disabled
06
Display mode
selection*
1
(2-axis models
only)
Selects the UNIT (counting method) that is assigned
to each CEL. The UNIT to be set can be selected as
desired.
For details, see "3.4.2 Details of Display Mode
Selection (Parameter Number 06)" (page 18).
● ● ● ●
07
Display at startup Selects stand-by state or Counter display (origin
detection wait state for EH-102Z) to display at startup.
● ● ● ●
EH-P/S/D
0: [--------] display
1: 0.000
EH-Z
0: [--------] display
1:
Origin detection wait state
13 µ
decimal point
display
(2-axis models
only)
If enabled, the decimal point will be displayed at the
µ
m position. This is available if the Linear Gage reso-
lution is 0.05
µ
m or less.
Example: 0.001.00 (1
µ
m)
● ● ● ●
0: Not displayed 1: Displayed
14
Sum/Difference
calculation
Species the calculation method when Sum/Differ-
ence calculation measurement (UNIT C) is selected in
parameter number 06. For 1-axis models, this is not
available.
● ● ● ●
0: A + B 1: A - B