3 Setting Parameters
16
No. 99MBC109B
No. Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
30
Analog output
range
Selects the range (resolution range) of the measure-
ment of the analog output.
0: 1999 to -1999
2: 199900 to -199900
1: 19990 to -19990
31
Origin detection
direction
When a Linear Gage with an origin mark is connected,
selects the direction of the spindle of the Linear Gage
for origin detection. It can be set per axis.
0: + direction 1: - direction
32
Origin re-detec-
tion
*3
When a Linear Gage with an origin mark is connected,
sets whether to wait for the origin to be detected with-
out turning off the power in the case of an abnormal
stop.
0: Disabled 1: Enabled
33
Origin initialization
(when the power
switch is on)
When a Linear Gage with an origin mark is connected,
initializes the origin when the power is on.
After the initialization, the set value will be returned to
0 (do not initialize).
0: Do not initialize 1: Initialize
*
1
The Preset value and tolerance value that had been set will be cleared if the setting is changed.
*
2
Turn off the power switch after changing the setting. The setting will be applied when you turn on the
power switch again.
*
3
When the setting is enabled, the Counter will wait for the origin re-detection when the HOLD signal is
raised. If the HOLD signal is input again during origin re-detection, the origin re-detection function will be
canceled (except during error detection).