Mitutoyo ID-C0550NXB - Page 81

Mitutoyo ID-C0550NXB
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
No. 99MAH057A
5 Setting Parameters
(Short press) / (Long press)
4
5
F2
F2
F2F2F2
F2 F2 F2
Refer to step
3
Refer to step
3
Go to the
next item
4 Set the calibration date.
1 Set in the same manner as the current date (step
3
).
2 Press the [F2] key.
» Advance warning date becomes settable and [AdV.] blinks.
5 Set the advance warning date.
1 Set in the same manner as the current date (step
3
).
2 Press the [F2] key.
» Settingisconrmed;shiftstothenextparameteritemof"Otherfunctionchanges".
(Go to "5.11.3 Selecting Digimatic Output" on page 70.)
Tips
If each date is set as follows, a calibration schedule setting error (Err 92) will occur.
Calibration date < Current date
Calibration date < Advance warning date
Advance warning date < Current date
Press the [F2] key and reset it so that current date < advance warning date <
calibration date. For details, see
"7 Error Displays and Countermeasures" on
page 75.