Tethys D60 - Tethys T60 - Tethys T45 SHORT-FORM MANUAL
2
1 SAFETY WARNINGS
1.1 KEY TO THE SYMBOLS USED
Read with particular attention (Symbol which appears beside particularly important instructions or warnings).
Warning, danger: refer to the manual. (The symbol appears on the appliance’s technical dataplate to
emphasise that staff must read the manual before using it. The symbol appears in the manual next to safety
instructions)
Warning, hot surface. (the symbol is affixed to the appliance close to parts which may heat to high
temperatures and constitute a burn hazard - do not touch parts near this symbol. The symbol appears in the
manual to emphasise safety instructions relating to the burn hazard).
Warning, electric shock hazard. (The symbol is affixed close to live parts - the appliance’s power supply must
be disconnected before doing any work on these parts. Never touch live parts unless the electricity supply is
disconnected).
Biological hazard.
DISPOSAL At the end of its working life, the product must be consigned to a disposal plant for recovery and
recycling in accordance with the relevant legislation in the country of installation. Contact the specialist
disposal consortia.This appliance is tagged under European Directive 2002/96/EC ON WASTE ELECTRICAL AND
ELECTRONIC EQUIPMENT (WEEE).
By making sure that this product is disposed of correctly, the user helps to prevent potential detrimental
effects on health and the environment.
CE marking; Notified body: IMQ. This symbol appears on the machine’s technical data label and in this
manual.
1.2 GENERAL WARNINGS
- Leaning or sitting on the open door of the appliance might cause it to tip over. The maximum weight which can be
loaded on the door, including the weight of the instrument trolley, must never exceed
Tethys D60, Tethys T60: 37kg
Tethys T45: 18kg
- if the weight of the instrument trolley and baskets is excluded, the maximum load must never exceed:
Tethys D60, Tethys T60: 23kg.
Tethys T45: 15Kg
- Only for models equipped with drying system: For optimal DRYING, the load for processing must never exceed 15 kg.
- In the event of malfunction, disconnect the appliance from the electricity supply and turn off the water tap. Then
contact your nearest authorised service centre.
- Open the door carefully, first waiting for the wash cycle to end.
- The appliance has an automatic door opening system; do not force the door open while a program is in progress.
- The appliance must only be used by staff suitably trained in its operation.
- The tank of the appliance is not designed for users to climb into it. The user must never climb into the tank; this might
put his safety at risk (ref. 7.102 IEC61010-2-040:2005).
Do not place flammable substances inside the appliance. Do not use flammable detergents.
Never place alcohol or solvents such as turpentine, which might cause an explosion, inside the appliance. Do not place
materials dirty with ash, wax or paint inside the appliance.
N.B. - ABSENCE OF DEMINERALISED WATER SUPPLY - If demineralised water is not available, the user is responsible for
ensuring that the quality of the water supplied to the appliance does not cause mineral salts or other substances to be
deposited on the instruments processed, rendering their subsequent use unsafe.
Never touch the heating element immediately after the end of a wash programme.
Do not touch any residual liquids left inside the washing chamber; scalding hazard (ref. 7.102.c IEC61010-2-040:2005).
When moving the appliance around, a forklift truck or pallet truck must be used.
Before leaving the factory, the base of the appliance is secured to a pallet, which is used for lifting and transporting it.
Do not use appliances which have been damaged in transit! If in doubt, contact your dealer.
CAUTION Warning applicable if the appliance is used for processing biologically contaminated material.
If a disinfection cycle is interrupted (by the user or due to an alarm generated by the appliance itself): take care when
handling the instruments and the load in general in the washing chamber. The load and the internal parts of the
appliance might be biologically contaminated/infected. Before handling the items in the load or performing any
maintenance procedures: perform a full thermal disinfection cycle or, if a complete cycle is not possible, handle the load
with due care (e.g. using the protective equipment specified when handling the infected load, e.g. gloves, lab coat).