What to do if Motorola Measuring Instruments has no horizontal sweep?
- CChris CastanedaSep 4, 2025
If your Motorola Measuring Instruments shows no horizontal sweep, check for a voltage level between -2.0 VDC and +2.0 VDC at TP4 of the Scope Amplifier module (A2). If the voltage cannot be brought within range with either the vertical range attenuator or the vertical position control, replace the Front Panel Interface module (A12). If the voltage at TP4 is okay, replace the Scope Amplifier module (A2).