SCREEN 2(5-6) SCREEN 3(5-6)
SCREEN 4(5-6) SCREEN 5(5-6)
B : Operation Instructions:
B-1 : Turn the rotary switch (3-13, Fig.1) to position
B-2 : When the input voltage signal, the system will automatically detect the correct
measurement frequency measurements made when the input voltage when
no signal, make sure the measurement frequency, at“4-5-14“option to select
the measurement frequency.
B-3 : The hook on the hook current signals, the LCD will display the waveform
and harmonic chart (eg SCREEN 2), ◄ or ► KEY choose H1 ~ H50 and
exhibits harmonic value, press ► KEY > 2 Sec then quickly switch to
H26 ~ H50 screen (eg SCREEN 3).
B-4 : Press FUNC KEY once the screen is switched to the harmonic map
features (as SCREEN 4), press ► KEY > 2 Sec then quickly switch to
H26 ~ H50 screen (as SCREEN 5), press it again to return to the SCREEN 2
5-7 Voltage and Current Waveforms Measured
1) Turn the rotary switch (3-13, Fig.1) to position
2) Connect the voltage and current signals, the LCD will display the voltage and
waveforms (as SCREEN 1)
3) Press FUNC KEY once the screen switches to SCREEN 2, press it again to
current return to the SCREEN 1.
25