5-8 Transient Measurement:
SCREEN 1(5-8) SCREEN 2(5-8)
SCREEN 3(5-8) SCREEN 4(5-8)
1) Turn the rotary switch (3-13, Fig.1) to position Transient
2) Connect the voltage signals, press REC KEY (3-6, Fig.1) event does not occur
(as SCREEN 1) screen ,when the Transient Voltage ≧ 242.0V LCD will display
SCREEN 2 screen (SWELL),upper right corner of the currently logged items.
when 40.0V ≦Transient Voltage≦ 198.0V when the LCD screen will display
SCREEN 3 (DIP),when the Transient Voltage≦ 40.0V when the LCD screen
will display SCREEN 4 (DIP-OUTAGE)
3) Press the REC KEY (3-6, Fig.1) and once released data logging mode,
press ▲ or ▼ KEY (3-8 or 3-11, Fig.1) to browse records in each state
(as SCREEN 2 ~ SCREEN 4).
5-9 Temperature Measurements:
1)Turn the rotary switch (3-13, Fig.1) to position Temp
2)Connect Type K Sensor, the screen will display SCREEN 1
3)℃ / ℉ switching, refer to 4-5-15. Their ℉ display screen such as SCREEN 2