EasyManua.ls Logo

MRC PC-6011SD - Transient Measurement; Temperature Measurements

MRC PC-6011SD
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCREEN 1(5-7) SCREEN 2(5-7)
5-8 Transient Measurement:
SCREEN 1(5-8) SCREEN 2(5-8)
SCREEN 3(5-8) SCREEN 4(5-8)
1) Turn the rotary switch (3-13, Fig.1) to position Transient
2) Connect the voltage signals, press REC KEY (3-6, Fig.1) event does not occur
(as SCREEN 1) screen ,when the Transient Voltage 242.0V LCD will display
SCREEN 2 screen (SWELL),upper right corner of the currently logged items.
when 40.0V Transient Voltage 198.0V when the LCD screen will display
SCREEN 3 (DIP),when the Transient Voltage 40.0V when the LCD screen
will display SCREEN 4 (DIP-OUTAGE)
3) Press the REC KEY (3-6, Fig.1) and once released data logging mode,
press or KEY (3-8 or 3-11, Fig.1) to browse records in each state
(as SCREEN 2 ~ SCREEN 4).
5-9 Temperature Measurements:
1)Turn the rotary switch (3-13, Fig.1) to position Temp
2)Connect Type K Sensor, the screen will display SCREEN 1
3) / switching, refer to 4-5-15. Their display screen such as SCREEN 2
26

Related product manuals