4 Start up
WARNING!
Installation should only be carried out by suitably qualified personnel.
Failure to follow this warning can result in serious personal injury or death.
Before applying power ensure that all installation steps have been correctly carried out and the cable connections and
controller configuration are correct.
CAUTION!
Ensure that the sensor current is correctly set to 310 mA for the 47K-PRP and 300 mA for the 47K-HT-PRP.
Failure to follow this caution can result in minor or moderate injury.
Allow a warm-up time of at least 30 minutes before calibrating the gas sensor.
Perform an initial calibration following the instructions in Section 5.1 Calibration and the manual for the relevant MSA
controller.
13 Series 47K GB