Sample Gas 100% LEL in
vol%
Propane relative
response factor
Response time
[secs] t
50
*)
Response time
[secs] t
90
*)]
Ethanol 3,1 0.97 < 14 < 31
1-Ethoxy-2 propanol
[30°C]
1,3 1.76 < 26 < 82
Ethyl acetate 2,0 1.19 < 16 < 36
FAM-Benzin 65/95
Gasoline 65/95
0,7 2.00 < 14 < 25
Hydrogen 4 0.59 < 9 < 17
Methane 4,4 0.58 < 11 < 18
Propane 1,7 1.00 < 12 < 19
1-Propanol [30 °C] 2,1 1.30 < 17 < 61
2-Propanol 2 1.07 < 15 <29
Propene 2 0.82 < 12 < 21
Toulene 1,0 1.36 < 21 < 54
n-Nonane [30 °C] 0,7 2.0 < 20 < 46
*) Response times are with a flow through adapter and a gas flow of 1 l/min
Response factor measured by DEKRA Testing and Certification GmbH
Other gases/vapours on request.