Maximum wind speed: up to 6 m/s
Applicable test gases:
H2, CH4, C3H8 in air (approval of other gases by MSA on
request)
Recommended flow rate: 1.0 l / min
Minimum flow rate: 0,8 l / min
Maximum flow rate: 1.5 l / min
Response times:
CH
4
t
50
15 s, t
90
40 s
C
3
H
8
t
50
20 s, t
90
55 s
For Part-No. of CalGard refer to Section 8.5 of this manual.
For mechanical installations of CalGard refer to Section 9.7 of this manual.
23 Series 47K GB