5.3.1 47K-HT-PRP
Relative response factors of tested gases with reference to Propane.
These values are only valid for new sensors and, unless otherwise stated, refer to
an ambient temperature of 110 °C. In this case the displayed values may vary by
up to ±20% from the target gas concentration.
Sample gas 100% LEL
in vol%
Propane relative
response factor
Response
time [secs]
t
50
*)
Response
time [secs]
t
90
*)]
Methane 4.4 0.52 < 8 < 14
*) Response times are with a flow through adapter and a gas flow of 1 l/min
5.3.2 47K-PRP
Relative response factors of tested gases with reference to Propane.
These values are only valid for new sensors and, unless otherwise stated, refer to
an ambient temperature of 20 °C. In this case the displayed values may vary by
up to ±20% from the target gas concentration. Response times stated in the
enclosed table may vary by ±20% depending on the tolerances of individual
sensors.
Sample gas
100%
LEL
in vol%
Propane relative
response factor
Response
time [secs]
t
50
*)
Response
time [secs]
t
90
*)]
Propane
1,7 1.00