PXIe-4466/4467 Calibration Procedure
36 ni.com/services
Table 27: Timebase Frequency Verification Configuration
Configuration Value
Physical Channels Dev_name/aix
Gain 0 (dB)
Max/Min Voltage +/-10V
Coupling AC
Terminal Configuration Pseudodifferential
Excitation (IEPE) 0 (A)
Acquisition mode Finite number of samples
Rate 40000.0
Samples per channel 2560000
4. Start the task.
5. Acquire reading with the DUT.
6. Measure the frequency using the Extract Single Tone Information VI of the
acquired signal around 10 kHz. Compare the measured frequency to the
limits in Table 27.
7. Stop and clear the task. Stop generating a signal with the function
generator.