PXIe-4466/4467 Calibration Procedure
4 ni.com/services
2 years
Password
NI
Note
This is the default password for all password-protected operations.
This password is site-specific.
Task Estimated Test Time
Operator
Connections
Test Points
Setup 10 minutes
Warm Up 15 minutes
Verify, Adjust, and
Reverify
19 minutes 36 344
Verify Only 8 minutes 17 171
Adjust Only 3 minutes 2 2
Note
Estimated test times assume the user is conducting a manual
calibration. For most procedures, automating the calibration
significantly reduces test times.
Environmental Conditions Verification Adjustment
Ambient temperature 23 °C ± 5 °C 23 °C ± 5 °C
Internal DUT temperature range Tcal ± 5 °C Tcal ± 5 °C
Relative humidity Between 20% and 80%, noncondensing