PXIe-4466/4467 Calibration Procedure
© National Instruments Corporation 7
Standard
Recommended
Model
Where Used
Functional
Requirement(s)
Calibrator Fluke 5730A AI Gain
AI Flatness
Frequency Range:
40 Hz to 100 kHz
Voltage Range: up to
6.3 V
rms
Digital Multimeter (DMM) Keysight (Agilent)
3458A
AO Differential Offset
AO Differential Offset
and Gain Adjustment
DC Voltage
Input Range: 100mV,
1 V, 10 V
AO Differential Gain
AO Differential
Flatness
AC Voltage
Input Range: 1 V
rms
,
10 V
rms
Frequency Range:
20Hz to 100 kHz
Function Generator Keysight (Agilent)
33250A series
Timebase Frequency Frequency Range: up
to 90 kHz
PXI Express Chassis NI PXIe-1082
NI PXIe-1085
NI PXIe-1075
NI PXIe-1065
NI PXIe-1062Q
All Tests
System Controller NI PXIe-8135 All Tests A PXI Express
controller or MXI
Express Card
mXLR (F) to BNC (M)
Cable (quantity 1)
NI 140150-0R46 All Tests (mXLR
Variant)
Length: ≤ 0.5m
BNC (M) to BNC (M)
Cable (quantity 1)
Pomona Electronics
5697
All Tests (BNC Variant) Length: ≤ 0.5m
BNC (F) to Banana
Adapter (quantity 1)
Pomona Electronics
1269
AI Gain
AI Flatness
AO Offset
AO Gain