19
Electric Connection
Electric installations of NeoQi® Harmony Pro
should be carried out by qualified specialists
or official representative of NeoQi® AS.
NeoQi® Harmony Pro is CLASS 1
EQUIPMENT in which protection against
electric shock does not rely on BASIC
INSULATION only, but which includes an
additional safety precaution, which implies
that the EQUIPMENT must be connected to a
protective earth conductor in a fixed wiring,
in such a way that accessible metal parts
cannot become live in the event of a failure
of the basic insulation.
NeoQi® Harmony Pro belongs to the group
of TYPE B EQUIPMENT, which provides a
degree of protection against electric shock,
which implies:
-allowable leakage current (10 mA);
-reliability of the protective earth connection.
NeoQi® Harmony Pro cannot be used if the
air contains FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR
NITROUS OXIDE.
Water Connections
The installation of warm and cold water
inlets and drainage connections of NeoQi®
Harmony equipment should be carried out
by qualified specialist or official
representative of NeoQi® AS.
For proper functioning of the equipment only
clean tap water should be used. In case the
tap water contains high percentage of
abrasive substances or rust, it is
recommended to install a water filter.