27
ERROR 3
Problem:
Displayed on the screen as E3. This alarm is
connected to the water level sensor error
(E2). If the signal E2 was present on the
screen for more than 2 minutes, the display
will signal E3. You will hear a sound: “Du,
Du, Du …” and control panel will not respond.
Solution:
Make sure that the steam generator is filling
up correctly. Check if the water supply is
connected to the device. Perform steam
generator cleaning - water level sensor may
be dirty and not be able to send a reading of
the level of water. If everything is well with
the water supply and there is no response,
check the steam generator data cable,
perhaps some of the contacts of the water
level sensor cables have eased.
ERROR 5
Problem:
Displayed on the screen as E5. It appears in
case the steam generator heating element is
overheating – if the temperature is over
108℃. When there is no water inside of the
steam generator, the heating element will
be turned off to protect it from overheating.
You will hear a sound: “Du, Du, Du …” and
the control panel will not respond.
Solution:
Stop the function, wait until the heater has
slightly cooled down. Check the water supply
to the steam generator, it might not be
filling correctly.
ERROR 6
Problem:
Displayed on the screen as E6. This occurs
when water cannot leave the steam
generator after the steam treatment. There
will be E6 on the screen, you will hear a
sound: “Du, Du, Du …” and the control panel
will not respond.
Solution:
Disconnect the device from main power
supply. Check the drain valve, it may be
stuck inside the socket or does not receive
the required voltage of 12V.