xiii
D.C. Voltage Measurement (Thermopile CAL MODULE)
Signal Range:
1, 2
01 2 3
Full Scale Voltage:
4
2.49 mV 24.9 mV 249 mV 2.49 V
(20,000 Count Precision)
Resolution: 125 nV 1.25 µV 12.5 µV 125 µV
(20,000 Count Precision)
Display Noise Floor: 8 LSD 1 LSD 1 LSD 1 LSD
(Input Shorted)
Full Scale Accuracy (Typ.):
3
± .3% ± .2% ± .1% ± .05%
Full Scale Accuracy: ± .56% ± .36% ± .18% ± .1%
(Worst Case)
Bandwidth (3db): 5 Hz 5 Hz 5 Hz 5 Hz
Pulse Voltage Measurement (Energy, Pyroelectric CAL MODULE)
Signal Range:
1
0123456789
Full Scale Voltage:
2,4
789 µV 2.50 mV 7.89 mV 25.0 mV 78.9 mV 250 mV 789 mV 2.50 V 7.91 V 25.0 V
A/D Resolution: 193 nV 610 nV 1.93 µV 6.10 µV 19.3 µV 61.0 µV 193 µV 610 µV 1.93 mV 6.10 mV
Full-Scale Accuracy:
3
±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1%
(Typical)
Full-Scale Accuracy:
3
± 2% ± 2% ± 2% ± 2% ± 2% ± 2% ± 2% ± 2% ± 2% ± 2%
(Worst Case)
Maximum Pulse Repetition Rate: 2 kHz
Trigger Level: 8% of Full Scale
1
(fixed)
1
Listed signal ranges specify meter capability. Available signal ranges are detector dependent.
2
Full scale voltage is measured relative to baseline voltage.
3
After 60 min warm-up, followed by AUTOCAL command. See Section 2.4.4.
4
Full scale voltage may vary due to AUTOCAL compensation of DC offsets.
Analog Output
Full Scale Voltage: 0 - 2.5V into 50
Accuracy: ± 5.0%