2.4.14 General Information Functions
General information about the Model 2835-C and any attached detector can be
found using the GENERAL INFO menu functions. These functions display the
Model 2835-C software version, the detector’s model number, serial number,
associated attenuator serial number and calibration date.
Table 20 – GENERAL INFO Operations.
Menu Function Keypad Commands Associated Remote Commands
MODEL cccccccc For viewing only. DETMODEL?
SW VERSION d.d For viewing only. *IDN?
DET SN ddddd For viewing only. DETSN?
ATTN SN ddddd For viewing only. ATTNSN?
CAL ddMonyyyy For viewing only. CALDATE?
2.5 Connecting AC Power
Model 2835-C can be configured to operate with line voltages over the
following ranges: 90-264 VAC, 50-60 Hz. Before turning the meter on, config-
ure it to local voltage using the following procedure:
i. Configure the Model 2835-C’s power supply voltage selection switches to
match the nominal local voltage. See Figure 5 and Table 21.
ii. Plug an AC line power cord to the rear of the Model 2835-C and then
connect the cord with AC power.
100 V
120 V
240 V
220 V
LINE
SELECT
50–60 Hz
SWITCH 1 SWITCH 2
Figure 5 – Rear Panel Power Supply Voltage Switches in Positions with
Switch 1 set to Left and Switch 2 set to Right
Table 21 – Power Supply Voltage Switch Positions.
Switch 1 Position Switch 2 Position Nominal Local Voltage
Right Right 100 VAC
Right Left 120 VAC
Left Right 220 VAC
Left Left 240 VAC