*RCL
Recall Meter Configuration
Syntax: *RCL <config>
Parameters:
<config> is of type <number> which rounds to an integer in the range 0 to 9.
This number specifies a configuration storage buffer. The number 0 stands for
the default configuration buffer.
Function:
This command configures the meter to a configuration previously stored in
non-volatile memory (see *SAV), or to the default configuration for the
attached detector. This command will fail if the requested configuration is
incompatible with the detector attached to the meter or empty. The param-
eters affected by this command are those listed in Table 9.
Returns: None
Related Commands: *RST, *SAV
*RST
Reset
Syntax: *RST
Parameters: None
Function:
This sets the meter to the default configuration, see Table 9, for the detector
attached to it. It also cancels any outstanding *OPC or *OPC? commands.
Related Commands: *RCL, *SAV
*SAV
Save Meter Configuration
Syntax: *SAV <config>
Parameters:
<config> is of type <number> which rounds to an integer in the range 1 to 9.
This number specifies a configuration storage buffer.
Function:
This command saves the configuration of the meter to the configuration
specified by the parameter. The parameters saved by this command are
those listed in Table 9.
Related Commands: *RCL, *RST