December 2019 72 LDC-3706 Series Laser Controller
CALibrate:PDI:STARt LAS
Description Begins the monitor photodiode current calibration procedure.
CALibrate:PDI:DATA <nrf> LAS
Description Send data point for the monitor photodiode current calibration procedure.
Parameters DATA – a number representing the actual current in Amps
Notes This is the same as CALibrate:MDI:STARt and CALibrate:MDI[:DATA] above; there are two versions for user
flexibility
Example CAL:PDI:DATA 0.0005; Sends a data point of 0.0005A.
CALibrate:RTD1000UA:STARt TEC
Description Begins the 1 mA RTD resistance calibration procedure.
CALibrate:RTD1000UA[:DATA] <nrf> TEC
Description Send data point for the 1 mA RTD resistance calibration procedure.
Parameters DATA – a number representing the actual resistance in Ohms
Example CAL:RTD1000UA:DATA 1000; Sends a data point of 1000.
CALibrate:RTD2500UA:STARt TEC
Description Begins the 2.5 mA RTD resistance calibration procedure.
CALibrate:RTD2500UA[:DATA] <nrf> TEC
Description Send data point for the 2.5 mA RTD resistance calibration procedure.
Parameters DATA – a number representing the actual resistance in Ohms
Example CAL:RTD2500UA:DATA 250; Sends a data point of 250.
CALibrate:SENSORI:STARt TEC
Description Execute the TE ICI sensor current calibration procedure.
Notes There must be nothing connected to the sensor pins.
CALibrate:SENSORV:STARt TEC
Description Begins the TE ICV sensor voltage calibration procedure.