EasyManua.ls Logo

Newport NewStep NSA12 - Page 55

Newport NewStep NSA12
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Set
55
PA move to absolute position
USAGE
IMM
MIP
SYNTAX
xxPAnn or xxPA?
PARAMETERS
Description
xx
[ int ]
-
Controller number
Nn
[ int ] - Position value
Range
xx
- 0 to 255
nn
- -60000 to 120000
Units
Micro-steps
Defaults
xx
Missing:
None
out of range:
None
nn
missing:
Error 37, Parameter Missing
out of range:
Error 7, Parameter Out of Range
Error 213, Motor Off
Error 230, Command not allowed
during motion
DESCRIPTION
This command initiates an absolute motion. When received, a
trapezoidal trajectory is generated with the predefined acceleration
and velocity and the actuator will move to the absolute position
specified by nn.
Note:
The command is not accepted while a motion is in progress.
RETURNS
If the “?” sign takes the place of nn value, this command reports the
last commanded position. Which is the same as the value read back
from TP? ± the predefined Backlash value, if there was a change of
direction.

Table of Contents