19
Research applications
Theresearchapplicationsarenotvalidatedfor
clinicaluse.
Special 6s FeNO measurement
Thisisaresearchapplicationwithanexhalation
timeof6seconds.
1. SelectMode
2. SelectSpecialmode
3. Makeameasurement
accordingtothe
instructionsonpage7,
withanexhalationtimeof
6seconds.
Nasal measurement
ContactyourlocalCircassiarepresentativeor
CircassiaTechnicalSupportforfurtherinformation.
Cautions
• Mobile phones and cordless phones might
interfere with the instrument and should
therefore be kept away from the instrument.
Interference could make it impossible to
perform a measurement.
• Make sure that you use the required
measurement mode, otherwise incorrect
exhaled NO results might be obtained.
•Useofsubstancescontainingalcoholcloseto
NIOX MINO
®
may cause erroneous measurement
results.
• Do not reuse the NIOX
®
Filters.
•TheNIOXMINOinstrumentmightproducesome
heatduringnormaloperation,thetemperature
couldincreaseupto5°Cabovetheambient
temperature.Makesurethatthebaseventilation
slotsarenotblocked.Donotplacetheinstrument
onabed,sofa,carpet,orothersoftsurface.
•Amaximumof10measurements/hourcanbe
performedduringcontinuoususe.However,it
willbepossibletoperform20measurements
perhouriftheinstrumentisallowedtorestfora
minimumof30minutesafterthat.
• TheSpecial6s.exhalationandnasalmodesare
intendedforresearchuseonly.
•TheNIOXMINOSensorshallbekeptinits
originalunopenedpackagebeforeinstallation.
Fortransportationandstorageconditions,referto
thecorrespondingsectioninthismanual.
•TheNIOXMINOSensorissensitivetochanges
inambienttemperatureandhumidity.
•BecarefulwhenopeningtheSensorcan.The
insideoftheopeninghassharpedges.Donot
touchorcleanthewhiteSensormembrane.
•Bestperformanceisachievediftheambient
conditionsarestable.Refertothe"Operating
conditions"section.Keeptheinstrumentaway
fromwindows,sunshine,radiators,stoveoropen
reinordertoavoidunstableconditions.When
transportingtheinstrumentfromonelocationto
another,aprolongedstabilizationperiodbefore
measurementmightberequired.Refertothe
recommended transportation conditions in the
“TransportandStorage”section.Alwaysusea
NIOXMINOBagfortransportation.Keepthe
NIOXMINOwithmountedSensorconnectedto
thepoweroutletwheneverpossible.
•NIOX MINO has been tested and found to comply
withthelimitsformedicaldevicesaccording
EN/IEC60601-1Safetyrequirementsfor
medicalelectricalsystemsandelectromagnetic
compatibility.Theselimitsaredesignedto
provideprotectionagainstharmfulinterference
inatypicalmedicalinstallation.However,
becauseoftheincreaseduseofradio-frequency
transmittingequipmentandothersourcesof
electricalnoiseemittersinthehealth-careand