2-8
Prime CCS Instructions for Use Manual
2.3.1 Manual Calibrations
A manually initiated 2-point calibration can be performed
whenever the analyzer displays Ready or Not Ready on
the header bar.
A Not Ready (Not Calibrated) status is displayed after
powering the analyzer on, after replacing some consumable
items or as a result of a system error. When the analyzer
displays Not Ready, samples cannot be run until a 2-point
calibration is performed that successfully calibrates the air
detectors and at least one analyte. To initiate a calibration
from the Not Ready state, press the Calibrate icon
on the Menu Bar.
A Ready status indicates the air detectors and one or more
analytes are calibrated and ready for analysis. To manually
calibrate the analyzer from the Ready state press Toolbox
icon then press Calibrate .
Analytes that display an orange background may be
uncalibrated. Press the icon and select Calibrate if
displayed to initiate a 2-point calibration.
Figure 2.6 Not Ready Screen and Ready Screen

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NOVA BIOMEDICAL Stat Profile PRIME and is the answer not in the manual?

NOVA BIOMEDICAL Stat Profile PRIME Specifications

General IconGeneral
BrandNOVA BIOMEDICAL
ModelStat Profile PRIME
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals