A-22
Prime CCS Instructions for Use Manual
Lactate Precision Data
Sample
Pooled
Mean
(mmol/L)
N
Within run
SD (Sr)
Within
run % CV
Total
imprecision
SD (St)
Total
Imprecision
%CV
QC Level 1 0.79 240 0.022 2.77 0.03 3.58
QC Level 2 2.60 240 0.012 0.47 0.01 0.54
QC Level 3 6.38 240 0.023 0.35 0.06 0.87
Linearity Std 1 13.36 240 0.229 1.71 0.79 5.91
Linearity Std 2 0.67 240 0.013 1.97 0.08 12.71
Linearity Std 3 2.53 240 0.022 0.87 0.07 2.88
Linearity Std 4 6.59 240 0.056 0.85 0.26 3.92
Linearity Std 5 10.51 240 0.086 0.82 0.45 4.26
A.3 Calibrator Cartridge
The concentrations of the internal standards are printed on
the Calibrator Cartridge. In addition to the calibrators and
solutions, the Calibrator Cartridge has a self-contained
waste bag for safe disposal of waste.
A.3.1 Traceability of Calibrators, Controls, and Standards
Chemistry analytes are traceable to the Standard Reference
Materials of the National Institute of Standards and
Technology (NIST). SO
2
is traceable to tonometry.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NOVA BIOMEDICAL Stat Profile PRIME and is the answer not in the manual?

NOVA BIOMEDICAL Stat Profile PRIME Specifications

General IconGeneral
BrandNOVA BIOMEDICAL
ModelStat Profile PRIME
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals