3-16
Prime CCS Instructions for Use Manual
If login is not required or after logging into the analyzer:
1.
From the Home Screen, press the QC button .
2. Press the Analyze QC button .
Figure 3.26 Quality Control Screens
3. From the drop-down list select the Prociency.
4. Press Start to begin the analysis.
5. Wait for the Sample Probe to fully extend.
6. Prepare the sample for analysis (mix well) then
position the sample
over the probe and
press Aspirate
. The sample
probe will retract
automatically once
sufficient sample
has been aspirated
into the analyzer.
Figure 3.27 Prociency Sample to Probe
7. Once the analysis is complete press Save to keep
the QC results or press Delete to discard the QC
results.