A-14
Prime CCS Instructions for Use Manual
Precision Study Results
Analytical Precision or Repeatability
Three Stat Prole Prime CCS Analyzers were assessed
for within run precision or repeatability within a laboratory
setting. The protocol consisted of within run precision, and
run–to-run precision studies using 3 different quality control
materials.
Stat Prole Prime Quality Control Level 1
Stat Prole Prime Quality Control Level 2
Stat Prole Prime Quality Control Level 3
Within Run Precision Performance
The protocol consisted of 20 replicates per run for each QC
level on each of 3 Stat Prole Prime CCS Analyzers. The
average, SD, CV%, and N for each analyzer for each QC
level and parameter was calculated. The pooled average,
SD, CV%, and N from all 3 analyzers for each QC level and
parameter was calculated.