4.2 UNPACKING
Remove the packing cardboard box and the second nylon packing around the centrifuge. The
below written are provided with the instrument, please check them;
1 ea. user’s manual
1ea. Warranty certification
1 ea. power cable
1 ea. manual lid opening tool
1 ea. 24 socket wrench
1 ea. Wrench
Lubricant oil
4.3 POSITIONING
Check that no damage has occurred during transport.
Check that the positioning is suitable for users.
Lift the centrifuge (use proper handling equipment if necessary) underneath and carry it to its
place. Check that the centrifuge is stable on its four pads.
The bench-top must be rigid enough to support the weight and vibrations.
Leave 30 cm. free space on the every side of the centrifuge.
Open the lid by using unlocking tool and check no substances left in the bowl. (See the manual
lid opening section 5.7 )
Make sure that the centrifuge does not occupy the utilization space of nearby equipment or do
damage to them.
The user should be able to follow the centrifuge during the operation.
According to the standard IEC 61010-2-020, anyone and any hazardous materials should not be
in the 300 mm safety zone while centrifuge is running.
ATTENTION !!!
The centrifuges are designed to operate safely under the following conditions:
Indoor use only
Ambient temperature: 5
°
C to 25
°
C.
Maximum relative humidity of 80% for temperature up to 22 °C.
Maximum altitude: 2000 m.
Temperature for maximum performance: 15 °C / 25 °C.
4.4 MAINS SUPPLY
The centrifuge requires 230 V, 50 / 60 Hz.
Please make sure that the supplied mains match the required power ratings. If no, provide an
extra line to support.
Always plug the sterilizer/ovens to properly earthed sockets.
A supply fitted with a circuit breaker should be used for protection against
indirect contact in case of an insulation fault.