8
Use only the the spare parts, rotors and accessories which are supplied by NUVE.
Load the rotor according to the explanations in the user’s manual.
Start the device after ensuring the rotor is loaded correctly.
Do not use the centrifuge in areas which are in explosive danger.
Do not centrifuge the explosive, flammable, radioactive, corrosive materials and the materials
which may react with each other.
The centrifuge and the rotor are not microbiologically leak-proof. Use tubes with leak-proof
covers, if hazardous, toxic and pathogenic microorganisms are centrifuged.
Do not use corrosive materials which may be harmful for the device integrity, rotor and
accessories.
Do not use rotors and accessories with corrosion and mechanical damages.
Mains supply should be appropriate to power of the device and grounded.
Use tubes whose sizes are suitable to the rotor and accessories.
Tubes which are used in the centrifuge should not be deformed by the effect of the centrifuge
force.
Use glass tubes to balance, if glass tubes are used. Use plastics tubes to balance, if plastics
tubes are used.
Do not start the device unless tubes are in balance.
Imbalance loading may cause mixing the samples, broken tubes, and damages on the rotor
and motor shaft.
IF MENTIONED WARNINGS ARE NOT CONSIDERED, NÜVE WILL NOT BE RESPONSIBLE
FROM THEIR RESULTS.
SECTION 3
SYMBOLS
Symbol in the operating instructions:
Attention, general hazard area.
This symbol refers to safety relevant warnings and indicates possibly dangerous
situations.
The non-adherence to these warnings can lead to material damage and injury to
personal.
Symbol in the operating instructions:
This symbol refers to important circumstances.
SECTION 4
INSTALLATION
4.1 LIFTING AND TRANSPORT
Because of the heavy weight of the centrifuge, all lifting and transport must be carried out using
proper handling equipment. The centrifuge should be lifted from underneath and never be turned
over.