NUVE NF 800 - Rotor Selection Table

Default Icon
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
LID
7
On / off switch
Lock pins
8
Bowl
Gas spring
9
Electrıcal cables inlet and fuses
Gasket
10
Manual LİD opening hole
Rotor
11
Cooling unit
Dısplay and conrol panel
2.4 ROTOR SELECTION TABLE
Rotor
Description
Capacity
Max. Tube
Rad
Max. Speed
Max RCF
Dim. Mm.
mm
Rpm
xg
RA 200
Swing-Out Rotor
4x 200 ml
Ø 57
162
4.100
3.044
Set of 4 buckets
200 ml
Ø 57
162
4.100
3.044
4 PP bottles with
cap
200 ml
Ø 56,5
162
4.100
3.044
Set of 4 inserts
1x100 ml
Ø 45
162
4.100
3.044
Set of 4 inserts
1x50 ml Conical
Ø 30
162
4.100
3.044
Set of 4 inserts
7x15 ml
Ø 17
162
4.100
3.044
Set of 4 inserts
3x15 ml Conical
Ø 17
162
4.100
3.044
Set of 4 inserts
7x5/7
Ø 13,5
162
4.100
3.007
MP 200
Microtitre Rotor
2 x 3 microtitre
125
4100
2349
RS 300
Angle Rotor
6 x 50 ml.
29.5
104
9000
9418
Set of 6 adapters
1 x 15 ml. conical
17
Set of 6 rubber
base adapters
1 x 50 ml. conical
29.5
RS 450
Angle rotor
30x15 ml
Ø 17
160
4.100
3.007
Adaptor
1 x 5 ml
Ø 13
160
4.100
3.007
Adaptor
1 x 7 ml
Ø 13
160
4.100
3.007
Adaptor
1 x 1.5/2 ml
Ø 11
160
4.100
3.007
RS 24
Angle Rotor
24x 1,5 / 2 ml
Ø 11
84
14.000
18.405
Set of 24 adapters
1x500 / 800µl
Ø 8
84
14.000
18.405
Set of 24 adapters
1x 200 µl PRC
Ø 6.5
84
14.000
18.405
Set of 24 adapters
1x250/400/700µl
Ø 6.5
84
14.000
18.405
Note: Insert set is not needed for 200 ml tubes.
2.5 PRECAUTIONS AND USAGE LIMITATIONS
Do not use the device for any purpose other than the usage purpose.
Prior to first use, the user’s manual should be read and the device is only to be used by
authorized and trained personnel. Only authorized technical personnel handle the product in
case of any failure.
The working bench should be durable to the device weight and vibration isolated.
Ensure that the rotor is placed correctly prior to usage.
According to the standard IEC 61010-2-020, anyone and any hazardous materials should not
be in the 300 mm safety zone while centrifuge is running.
Do not move the device while it is running.
Do not open the lid while rotor is spinning.
Apply the manual lid opening procedure in the case of power cut or in the case of any error.