12 Practical Guide for operating the C-Quant
42 / 56 Instruction Manual C-Quant ( G/80000/1007/en)
Example 3a: esd > 0.12 in “G” range
If there is no time to repeat the measurement, this measurement can be used to es-
timate an upper limit for the straylight value.
Because there is a consistent row of 1 responses in the right part of the graph, it is
quite certain that the patient does not have a straylight value higher than the out-
come of the measurement.
If you need a more reliable straylight value, just repeat the measurement.