12 Practical Guide for operating the C-Quant
44 / 56 Instruction Manual C-Quant ( G/80000/1007/en)
Example 3c: esd > 0.12 in “G” range
This example is similar to Examples 3a and 3b, only esd is so high that no estimation
for the straylight value could be calculated (log(s)=6.00 is a meaningless value).
However, you can make this estimation yourself by looking at the 1 and 0 responses:
the 50%-point of the curve will probably be not higher than 1.8, corresponding to a
straylight value of 1.5, a moderately increased value.