4 Safety Instructions
8 / 85 Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
Patient environment information
Patient environment is the area where patients can come into contact
with any part of a medical electrical equipment (ME equipment) or with
another person being in contact with the ME equipment.
Attention
In the patient environment, use devices that conform to IEC 60601-1. If
a multiple power socket is to be used, or if a device is to be used that does
not meet the IEC 60601-1 standard, use an isolating transformer.
Information about the operation of an ME system
The Pentacam® AXL and a connected computer form a medical
electrical system (ME system) according to IEC 60601-1. If you connect
Fig. 4-1: Patient environment