4 Safety Instructions
10 / 85 Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
Instructions for Maintenance
In order to retain the high measurement accuracy of the Pentacam® AXL
OCULUS Optikgeräte GmbH recommends to perform a maintenance
service every two years or after 25 000 scans. A corresponding message
appears, see sec. 14.3, page 65. Additionally to that it is usefull to
accomplish a test measurement of the axial length measuring mode
everyday before you start working with the Pentacam® AXL.
If an error occurs which you cannot correct, label the Pentacam® AXL as
being "out-of-order" and contact our service department, see sec. 18,
page 70.
Instructions for Disassembly and Disposal
When disconnecting electrical connections, pull on the respective
plug and not on the cable itself.
Dispose of the device according to legal regulations.
Instructions on Electrical Safety
Attention
Risk of personal injury or damage to property due to an incorrect level of
safety
Connecting the Pentacam® AXL with its non-medical electrical
equipment (e.g. data processing equipment) to a medical electrical
system must not result in a patient safety level below that prescribed by
IEC 60601-1. If making this connection leads to the leakage current
threshold being exceeded, protective measures including a circuit breaker
must be in place.
Ensure that connections with non-medical devices are made
correctly.
Only use the power adapter listed in the packing list.
Use only a computer that meets the specifications given in this
instruction manual, sec. 19, page 72.
Use of a multiple socket extension cord
Risk of personal injury or material damage caused by unsafe multiple
socket extension cord
If you use a multiple socket extension cord to connect the
Pentacam® AXL to the power supply, you must heed the following
information:
Use an extension cord that complies with the requirements of IEC
60601-1, section 16.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Pentacam AXL and is the answer not in the manual?

OCULUS Pentacam AXL Specifications

General IconGeneral
BrandOCULUS
ModelPentacam AXL
CategoryMedical Equipment
LanguageEnglish

Related product manuals