14 Cleaning, Disinfection and Maintenance
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04) 63 / 85
Cleaning intervalls
Clean the components of the Pentacam® AXL once a month or if
necessary.
Cleaning the Housing
It is best to clean the housing surfaces with a soft cloth and an
anti-static cleaning agent.
Wipe off any residue from painted surfaces with the mixture for
painted surfaces.
Cleaning the Chin-Forehead Rest
Make sure that no liquid gets into any of the openings of the
Pentacam® AXL.
Do not use liquid from aerosol cans.
Clean the chin-forehead rest with a soap solution (or with alcohol,
if very dirty).
Use a lint-free, damp cloth.
Cleaning the illuminated slit
The optical components illuminating the slit, and the lens in front of the
camera are precision parts and sensitive to pressure. The surfaces of these
components are susceptible to scratching.
Attention
When cleaning the illuminated slit, do not use a cloth or other cleaning
agents.
Carefully clean the centre of the illuminated slit with purified
compressed air.
Clean the lens in front of the camera using a dry, lint-free cloth.
Clean the acrylic glass with a commercial clean agent.