PR Series Balance EN-35
Table 9-3 SPECIFICATIONS (continued)
Verification interval e(g)
Repeatability (STDEV) (g)
Not exceeding the maximum permissible error for the one-third of the full capacity of the
balance
Stabilization Time Typical (s)
Sensitivity Temperature Drift
(PPM/K)
±3 ±3 ±3 ±3
Typical Minimum Weight USP
(USP K=2,U=0.10%)
200mg 200mg 2g 2g
Optimized Min-Weight
(USP, u=0.10%,
k=2)
82mg 82mg 0.82g 0.82g
g, mg, ct, grain, oz, ozt
Basic Weighing; Parts counting; Percent weighing
Span Calibration Points (g)
Linearity Calibration Points (g)
50, 75, 100
100, 200, 300
To capacity by subtraction
Power input: 100-240 V ~ 200 mA 50-60Hz 12-18VA
Power output: 12 VDC 0.5A
Assembled Dimensions
(W x D x H)
201 x 317 x 303 mm
7.9 x 12.5 x 11.9 inch
Operating Temperature Range
Storage Temperature Range
Humidity: maximum relative humidity 80 % for temperatures up to 30°C, decreasing
linearly to 50% relative humidity at 40°C
-10°C to 60°C, humidity 10% to 90%, without condensation
Shipping Dimensions
(W x D x H)
507 x 387 x 531mm
20 x 15 x 21 inch
*SRP refers to the standard deviation for n replicate weighings (nŐ10).