STARTER 3100C Bench Conductivity Meter EN-9
Use or buttons to adjust the temperature correction value. (0.00 to 10.00 %/°C).
Press to confirm your setting. Continue with reference temperature setting by following
or press to go back to the measurement screen.
Note:
If you set the temperature correction coefficient value as 0 %/°C , that means NO Temperature
Compensation for the conductivity measurement. The meter displays the real conductivity value
at current temperature.
3.5 Set TDS factor
After selecting the reference temperature, the current TDS factor blinks.
Use or buttons to adjust the TDS factor value. Press to confirm your
setting or press to go back to the measurement screen.