STARTER 3100C Bench Conductivity Meter EN-5
2 INSTALLATION
Carefully unpack the meter.
2.1 Package contents
The ST3100C-B package has the following items
In addition to ST3100C-B content, the ST3100C-F package also includes the following:
The best measurement range for STCON3 4-ring (4-pole) conductivity electrode is 70 μS/cm-200
mS/cm. In this range the measurement accuracy can reach 0.5%.
If you use STCON3 to measure pure water in which conductivity value is below 70 μS/cm; the
measurement will be less accurate (approximately 1% to 5%).
Does not use STCON3 to measure cond. value less than 2 μS/cm, the accuracy is not good.
Accessories:
Model
Description
P/N
STCON3
4-ring cond. electrode
(70 μS/cm-200mS/cm, 0.5% accuracy; )
( 2-70 μS/cm, 1-5% accuracy)
83033972
STCON7
Pure water cond. electrode
(0.02μS/cm-200 μS /cm; 0.02 μS/cm accuracy)
30080693
Standard Conduct 10 μS/cm 250ml
30100441
Standard Conduct 84 μS/cm 250ml
30100442
Standard Conduct 1413 μS/cm 250ml
30100443
Standard Conduct12.88mS/cm 250ml
30100444
Electrode holder stand alone (3100/3100C)
30058733
In use cover
30058734
ST3100C-B
Units
STARTER 3100C meter
1
Stand alone electrode holder
1
1413 µS/cm standard solution
1 bottle (about 20ml)
12.88 mS/cm standard solution
1 bottle (about 20ml)
In use cover
1
STCON3 conductivity electrode
1

Table of Contents