6 Model 5380 PFPD Operator’s Manual
Rev. 3.1
Constant reignition minimizes “flameout” problems, common in standard
FPDs, that are caused by water and other solvents.
WinPulse software provides real-time display of the emission resulting from
each flame pulse. It facilitates detector setup and configuration, optimization
of operating parameters, and troubleshooting without the use of an oscillo-
scope.
Post-run processing of data through the use of optional PFPDView
software.
Specifications
Detector Linearity
Sulfur—Quadratic in response. Linear over 2.5 orders of concentration, which
gives five orders of signal response.
Phosphorus—First order linear over five orders of magnitude.
Detectivity
Sulfur: < 1 pg S/sec
Phosphorus: < 100 fg P/sec
Selectivity at Optimum Detectivity Levels
Sulfur: > 10
6
S/C (or greater depending on gate settings)
Phosphorus: > 10
6
P/C
(Selectivity is adjustable with a trade-off in detectivity.)
Column Flow Range
Optimum performance: 1–3 mL/min
Maximum: 5–6 mL/min
Drift
Sulfur or Phosphorus: < 10x peak-to-peak noise in 20 min
Temperature Limitations
Minimum: 180°C
Maximum: 420°C
Heating Mode
Direct heating by heater cartridge
RTD (PRT) or thermocouple temperature sensing
Environmental Considerations:
Humidity: 5%–80% relative humidity
Temperature: 10°–40°C (Operating)
-20°– 65°C (Nonoperating)
Altitude: Maximum 2,000 m
Controller Dimensions
8.75" H x 6" W x 13" D
22.2 cm H x 15.2 cm W x 33 cm D