7 - Technical Specifications
63
Table no. 3
Carbon dioxide
(CO
2
)
Carbon Monoxide
(CO)
Hydrogen sulfide
(H
2
S)
Sensor P/N
6313818
6313787
6313788
Standard range (1)
0 - 5% v/v
1000
100
Measurement principle
Absorption
Infrared
Electrochemical
Electrochemical
Display resolution(1)
0.1% v/v
1
1
Accuracy (2)
0.2% v/v
15
3
Repeatability (3)
0.1% v/v
1
1
Zero/Span drifty (4)
0.2 / 2
0.5 / 1.5
0.5 / 2.5
Response time (5)
< 30s
< 30s
< 25s
Temperature (6)
-10°C to +40°C
-20°C to +40°C
-20°C to +40°C
Relative humidity and
pressure range (7)
10 95% RH
bar ± 20%
10 95 % RH
1 bar ± 20 %
10 95% RH
1 bar ± 20%
Lifetime (8)
60 months
36 months
36 months
Storage conditions and
maximum storage time
(9)
0°C 40°C
10 60% RH
1 bar ± 20%
6 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
Warm-up Delay
120s
Operational upon start up
Notes
Exposure to high levels of organic solvents
can damage the sensors.
Exposure to gases at levels higher than the
monitor’s range can damage the device.
Recalibrate sensors if they go out of range.