64
BM 25/25W
Instruction Manual
Table No. 4
Chlorine
(Cl
2
)
Hydrochloric acid
(HCl)
Hydrogen cyanide
(HCN)
Sensor P/N
6313809
6313804
6313805
Standard range (1)
10
30
10
Measurement principle
Electrochemical
Electrochemical
Electrochemical
Display resolution(1)
0.1
0.1
0.1
Accuracy (2)
0.25
1
0.25
Repeatability (3)
2
2
2
Zero/Span ndrift (4)
0.5 / 5
0.5 / 5
0.5 / 5
Response time (5)
< 60s
< 80s
< 60s
Temperature (6)
-20°C to +40°C
-20°C to +40°C
-20°C to +40°C
Relative humidity and
pressure range (7)
10 90% RH
bar ± 20%
15 95% RH
1 bar ± 20%
15 95% RH
1 bar ± 20%
Lifetime (8)
30 months
24 months
24 months
Storage conditions and
maximum storage time
(9)
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
Warm-up Delay (10)
Operational upon start up
Notes
Exposure to high levels of organic solvents can damage the sensors.
Exposure to gases at levels higher than the monitor’s range can
damage the device. Recalibrate sensors if they go out of range.