7 - Technical Specifications
65
Table no. 5
Ammonia
(NH
3
)
Ammonia
(NH
3
)
Nitrogen oxide
(NO)
Sensor P/N
6313799
6313800
6313802
Standard range (1)
100
1000
300
Measurement principle
Electrochemical
Electrochemical
Electrochemical
Display resolution(1)
1
1
1
Accuracy (2)
5
30
10
Repeatability (3)
2
2
1
Zero/Span drift (4)
1 / 2
1 / 2
0.5 / 3
Response time (5)
< 90s
< 90s
< 30s
Temperature (6)
-20°C to +40°C
-20°C to +40°C
-15°C to +40°C
Relative humidity and
pressure range (7)
10 90% RH
bar ± 20%
15 95% RH
1 bar ± 20%
15 95% RH
1 bar ± 20%
Lifetime (8)
30 months
24 months
30 months
Storage conditions and
maximum storage time
(9)
4°C 20°C
10 60 % RH
1 bar ± 10%
2 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
4°C 20°C
10 60% RH
1 bar ± 10%
2 months
Warm-up Delay (10)
Operational upon start up
Notes
Exposure to high levels of organic solvents can damage the sensors.
Exposure to gases at levels higher than the monitor’s range can
damage the device. Recalibrate sensors if they go out of range.