EasyManua.ls Logo

Olympus BH2 Series - Page 10

Olympus BH2 Series
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A.
TROUBLESHOOTING (B2-BDT)
Replace
the
QlOl.
Is
the
lamp
normally
lit?
YES
~~N-0-----1B
Replace
the
UYPC48.
2-4
Lamp
burns
out.
S\J
101
S\J
102
F101
~
channe~
T101
Causes
No.
A-6
UVPC4B
l.
Defective
transistor
QlOl
2.
Defective
UYPC48
@
Is
the
transistor
QlOl
normal?
NO
l
YES
0---------,
t
I
Replace
the
QlOl.
I
Is
the
lamp
lit?
I
YES
-t
~NO
I
Replace
the
UYPC48. J
G0
Normal
if
0.6V
or
greater
voltage
is
generated
between
the
base
(yellow)
and
the
emitter
(brown)
of
QlOl
when
measured
with
a
multimeter
with
the
light
intensity
control
lever
set
to
the
Max.
position.
Abnormal
if
the
voltage
is
lower
than
0.6V.

Other manuals for Olympus BH2 Series