EasyManua.ls Logo

Olympus BH2 Series - Display; 3-L Abnormal Display

Olympus BH2 Series
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B.
TROUBLESHOOTING
(B2-BDS-2)
No.
B-5
3.
Display
3-l
Abnormal
display
SW!03
T 1 0 1
LAMP
J3J2
UYPC65
JJ03
J201
JJ02
Causes
l.
Defective
variable
resistor
RVlOl
2.
Defective
UYPC65
3.
Defective
UYPC64
4.
Defective
power
supply
part
*
The
display
with
LED
is
unstable
when
a
lamp
is
burnt
out
or
not
mounted.
Be
sure
to
mount
a
lamp
and
set
the
primary
voltage
selector
switch
SW103
(the
100/llSV
selector
switch
in
the
lOOV
channel
and
the
220/240V
selector
switch
in
the
200V
channel)
to
the
appropriate
position.
Is
the
variable
resistor
RVlOl
normal?
YES
NO
Replace
the
RVlOl.
Does
the
display
with
LED
change
with
the
adjustment
of
the
light
intensity
control?
YES
.
~
NO
I
@
Measure
the
resistance
between
[l]
(brown)
and
@]
(orange)
of
J201
with
a
multimeter.
Normal
if
the
resistance
value
varies
smoothly
from
0
to
2k~
when
the
light
intensity
control
lever
is
slid.
Abnormal
if
not.
~
'
Proceed
the
work
according
to
"No.
J
-7
4,
Power
Supply,
[]] "

Other manuals for Olympus BH2 Series