EasyManua.ls Logo

Olympus BH2 Series - Power Is Not Supplied Even after the Main Switch Is Turned on; Power Supply

Olympus BH2 Series
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B.
TROUBLESHOOTING
(B2-BDS-2)
No.
B-6
4.
Power
Supply
4-1
Power
is
not
supplied
even
after
the
main
switch
is
turned
on.
SWI03
TIOt
LAMP
UYPC65
~-----J_2_o_tJ:~~~=:=:=:=:=:=:=:=:=:=:=:~====================~
Jt02
Causes
1.
The
breaker
functions.
2.
Blown
main
fuse
FlOl
3.
Defective
UYPC65
4.
Defective
UYPC64
5.
Defective
transformer
Tl02
6.
Defective
transformer
TlOl
7.
Defective
main
switch
SW102
8.
Defective
primary
voltage
selector
switch
SW103
9.
Defective
breaker
SWlOl
10.
Defective
noise
filter
RClOl
11.
Defective
spark
gap
RC102/RC103
12.
Defective
AC
inlet
JlOl
13.
Defective
AC
cord
(UYCPlS
in
lOOV
channel)
Is
the
breaker
functioning?
NO
YES
0-----c~~~-®
Press
the
breaker
button.
Is
power
supplied
normally?

Other manuals for Olympus BH2 Series